Gain and Read Noise

The conversion gain of the camera and the readnoise was measured by the following procedure: Note: outliers outside of 5 sigma in the difference images were rejected to calculate the variance.

Note: This value differs from the value of 30 derived for the Mark I camera by more than the expected errors. I do not know the reasonf for this discrepancy as the electronics have not been changed.

Gain Discussion

We can convert the conversion gain into the signal by following the signal chain which has the following characteristics: Hence 1 e- corresponds to about 1 microvolt. This is a factor of two lower than mean value of the range (1.5 - 2.5 µ V/e -) quoted on the Rockwell Specification Sheet. I don't know if this indicates some problem in the measurement,however, there are other reasons to be suspicious of this value.

Noise Discussion

The noise resulting from the difference of two Fowler frames is pretty white as the next image shows.


[ IRCam Index ]

Last Revised November 2, 2000